Umano ATP13A2 Kit Elisa (A74668)

Questo prodotto è fuori produzione

Umano ATP13A2 Kit Elisa (A74668) è stato interrotto e non è più disponibile.

Visualizza tutti gli prodotti disponibili.

Nome
Human ATP13A2 ELISA Kit
Tipo di kit
Sandwich (quantitative)
Principio di kit
This Human ATP13A2 ELISA Kit is an in vitro competitive enzyme-linked immunosorbent assay for the quantitative measurement of ATP13A2 in plasma, tissue homogenates and other biological fluids.

A microplate has been precoated with Anti-ATP13A2 Antibodies. Samples and the ATP13A2-HRP conjugate are added to the wells, where ATP13A2 in the sample competes with the added ATP13A2-HRP for antibody binding. After incubation, the wells are washed to remove unbound material and TMB substrate is added. The TMB substrate is catalyzed by HRP to produce blue colouration. The reaction is terminated by addition of Stop Solution which stops the colour development and produces a colour change from blue to yellow. The intensity of the signal is inversely proportional to the amount of ATP13A2 in the sample and the intensity is measured at 45 nm.
Piattaforma
Pre-coated Microplate (12 x 8 well strips)
Tipo di rilevamento
Colorimetric
Instrument
Colorimetric Microplate Reader
Tipo di campione
Serum, plasma, tissue homogenates, and other biological fluids.
Sensibilità
0.188 ng/ml
Campo di rilevamento
0.313-20 ng/ml
Tempo del kit
4h 30m
Reattività
Human
Recovery
Sample Type n Range Average
Serum 5 89% - 101% 95%
EDTA Plasma 5 89% - 98% 92%
Heparin Plasma 5 89% - 102% 95%
Linearity
Sample Type n 1:2 1:4 1:8
Serum 5 85-99% 86-103% 85-101%
EDTA Plasma 5 82-99% 82-94% 83-93%
Heparin Plasma 5 81-97% 83-97% 87-98%
Precision
Intra-assay CV: < 8%, Inter-assay CV: < 10%
Note generali
Information online is representative. Refer to the Instructions for Use inside the kit.
Conservazione
Store at +4°C. Do not use past expiration date!
Dichiarazione di non responsabilità
Questo prodotto è solo per uso di ricerca. Non è inteso per uso diagnostico o terapeutico.
Inizio pagina