Mouse IL-6 ELISA Kit is a 3 hour indirect Enzyme-Linked Immunosorbent Assay (iELISA) designed for the in vitro quantitative determination of mouse IL-6 in serum, plasma, cell culture supernatant, cell or tissue lysate, and other biological fluids.
Assay Type
Indirect (quantitative)
Principle of Assay
Mouse IL-6 ELISA Kit (A350374) is a 3 hour indirect Enzyme-Linked Immunosorbent Assay (iELISA) designed for the in vitro quantitative determination of mouse IL-6 in serum, plasma, cell culture supernatant, cell or tissue lysate, and other biological fluids. A 96-well microtiter plate has been pre-coated with assay-specific antigen. Standards and samples are added to the wells, where IL-6 binds to the immobilized antigen. Following incubation, unbound components are removed by washing and an enzyme-conjugated detection antibody is added to the wells. This detection antibody binds the captured IL-6 to form an indirect immune complex. After further incubation and washing, TMB substrate solution is added to visualize the enzyme-mediated reaction. This produces a blue-coloured product that changes to yellow following addition of acidic stop solution. The intensity of the yellow colour is directly proportional to the amount of IL-6 present in the sample. The concentration of IL-6 is calculated by reading the absorbance at 450 nm and comparing the values with the standard curve.
Platform
Pre-coated Microplate (12 x 8 well strips)
Detection Type
Colorimetric Detection (OD450)
Instrument
Colorimetric Microplate Reader
Sample Type
Serum, plasma, cell culture supernatant, cell or tissue lysate, and other biological fluids.
Sensitivity
0.469 ng/mL
Product Range
0.781-50 ng/mL
Assay Time
3 hours
Reactivity
Mouse
Cross Reactivity
Specific for IL-6 with no reported cross-reactivity with other analogs or variants.
Recovery
Sample Type
n
Range
Average
Serum
10
89% - 104%
98%
EDTA Plasma
10
92% - 103%
93%
Heparin Plasma
10
86% - 105%
93%
Linearity
Sample Type
n
1:2
1:4
1:8
Serum
10
85-95%
82-100%
84-96%
EDTA Plasma
10
86-96%
85-104%
92-102%
Heparin Plasma
10
84-99%
88-93%
90-100%
Storage
Shipped at +4°C. Store kit components at +4°C or -20°C as instructed in the Protocol Manual. Do not use past expiration date! Avoid freeze/thaw cycles.
Linearity assessment of the Mouse IL-6 ELISA Kit (A350374) across serial sample dilutions. Linearity was evaluated using Serum, EDTA Plasma, and Heparin Plasma samples serially diluted from 1:2 to 1:8. Recovery (%) was calculated relative to the expected Mouse IL-6 concentration at each dilution. Data are presented as mean recovery values from 10 replicates (n = 10).
Publishing research using Mouse IL-6 ELISA Kit (A350374)? Please let us know so that we can list the citation on this page.