Principle of Assay
Human CD11a ELISA Kit (High Range) (A319386) is a 4 hour sandwich Enzyme-Linked Immunosorbent Assay (sELISA) designed for the in vitro quantitative determination of human CD11a in serum, plasma, tissue homogenates, and other biological fluids. A 96-well microtiter plate has been pre-coated with an antibody specific for CD11a. Standards and samples are added to the wells, where CD11a binds to the immobilized capture antibody. Following incubation, unbound components are removed by washing and a biotinylated detection antibody specific for CD11a is added to the wells. The detection antibody binds the captured CD11a to form an antibody-antigen-antibody sandwich complex. After further incubation and washing, HRP-Streptavidin conjugate is added to the wells. TMB substrate solution is then added to visualize the HRP enzymatic reaction. This produces a blue-coloured product that changes to yellow following addition of acidic stop solution. The intensity of the yellow colour is directly proportional to the amount of CD11a captured in each well. The concentration of CD11a is calculated by reading the absorbance at 450 nm and comparing the values with the standard curve.