Principle of Assay
Human Phosphorylated Tau ELISA Kit (A74015) is a 4 hour sandwich Enzyme-Linked Immunosorbent Assay (sELISA) designed for the in vitro quantitative determination of human Phosphorylated Tau in serum, plasma, tissue homogenates, and other biological fluids. A 96-well microtiter plate has been pre-coated with an antibody specific for Phosphorylated Tau. Standards and samples are added to the wells, where Phosphorylated Tau binds to the immobilized capture antibody. Following incubation, unbound components are removed by washing and a biotinylated detection antibody specific for Phosphorylated Tau is added to the wells. The detection antibody binds the captured Phosphorylated Tau to form an antibody-antigen-antibody sandwich complex. After further incubation and washing, HRP-Streptavidin conjugate is added to the wells. TMB substrate solution is then added to visualize the HRP enzymatic reaction. This produces a blue-coloured product that changes to yellow following addition of acidic stop solution. The intensity of the yellow colour is directly proportional to the amount of Phosphorylated Tau captured in each well. The concentration of Phosphorylated Tau is calculated by reading the absorbance at 450 nm and comparing the values with the standard curve.